NSN 6625-01-650-1858
Part Details | CURRENT-VOLTAGE STANDARD
6625-01-650-1858 An item designed to produce an alternating current, direct current, and electromotive force with high degree of accuracy. It serves as a standard of calibration for AMMETER(S), VOLTMETER(S), WATTMETER(S), and similar devices that require accurate voltage and/or current inputs for calibration purposes. See also VOLTAGE STANDARD. For items whose outputs are used to operate other electrical or electronic equipment, see POWER SUPPLY.
Alternate Parts: VI700, VI-700, 6625-01-650-1858, 01-650-1858, 6625016501858, 016501858
| Supply Group (FSG) | NSN Assigned | NIIN | Item Name Code (INC) |
|---|---|---|---|
| 66 | NOV 23, 2015 | 01-650-1858 | 22982 ( CURRENT-VOLTAGE STANDARD ) |
REFERENCE DRAWINGS & PICTURES
Cross Reference | NSN 6625-01-650-1858
| Part Number | Cage Code | Manufacturer |
|---|---|---|
| VI-700 | 088V6 | IET LABS INC |
Technical Data | NSN 6625-01-650-1858
| Characteristic | Specifications |
|---|---|
| FREQUENCY RATING | 50.0 HERTZ MINIMUM AND 60.0 HERTZ MAXIMUM |
| OUTPUT VOLTAGE RATING | 0.0 VOLTS MINIMUM AND 200.0 VOLTS MAXIMUM |
| OUTPUT CURRENT RATING | 200.0 AMPERES NOMINAL |
| OVERALL DIAMETER | 9.25 INCHES NOMINAL |
| OVERALL WIDTH | 8.5 INCHES NOMINAL |
| OVERALL HEIGHT | 4.44 INCHES NOMINAL |
| PORTABILITY | PORTABLE |
| SPECIAL FEATURES | OPERATING TEMPERATURE RANGE: -20 DEGREES C TO 40 DEGREES C AND -20 DEGREES C TO 28 DEGREES FOR HP-HIGH POWER OPTION; POWER REQUIREMENT 105-125 V OR 210 - 250 V; 5 WATTS |
| PART NAME ASSIGNED BY CONTROLLING AGENCY | VI-700 MANUAL OR PROGRAMMABLE VOLTAGE AND CURRENT SOURCE |